Abstract
A small low-temperature plasma (LTP) ionization probe was coupled to a portable mass spectrometer for the rapid detection of trace explosives on surfaces. Using only a small diaphragm pump to supply ambient air to the LTP source, 100 ng each of pentaerythritol tetranitrate (PETN), 1,3,5-trinitroperhydro-1,3,5-triazine (RDX), and 2,4,6-trinitrophenylmethylnitramine (Tetryl) were detectable on glass in under 1 minute. The main ion signal from these molecules (M) is the [M + NO3]− species. While much optimization remains, it is believed that this miniature LTP source will remove the need for external gas cylinders and additional heating for in situ explosives detection using portable mass spectrometers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.