Abstract

To solve the problem of temperature drift of thermocouples in accelerating rate calorimeter caused by long-term operation, an in-situ calibration method based on multiple fixed-points and Joule heat is proposed in this article. Firstly, the heat transfer model of the calorimeter is established, and the validity of the method is verified by numerical simulation. Secondly, a multiple fixed-points graphite calibrator and a cylindrical electronic resistance element are designed. Finally, the in-situ calibration is carried out. The calibration results show that the maximum permissible measurement error of the sample thermocouple after calibration is better than 0.220 °C and that the bias of consistency in-situ calibration method is smaller than 0.110 °C. In addition, a di‑tert‑butyl peroxide in toluene solution with a mass percent of 20% is selected as the experimental sample. The sample experiment results show that the kinetic and thermodynamic parameters are closer to the reference values after thermocouples calibration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.