Abstract

The indiscriminate use of antibiotics in dairy cattle without complying with the waiting period results in residual contamination, whose effective control in produced milk requires validated methods toensure analytical results. The aim of this study was to optimize and validate the HPLC-UV/VIS method at 365 nm for analyzingthe tetracycline in pasteurized cow milk in accordance with the European Community (2002/657/EC). Spiked milk with analytes (oxytetracycline, tetracycline, doxycycline, and chlortetracycline) was submitted to deproteinization and cleaning by a C18 solid-phase column and analyzed by HPLC using a gradient system with 0.01 mol L?1 oxalic acid-acetonitrile-triethylamine (90:9.9:0.1) and acetonitrile on a reverse phase (C18) column. Accuracy and precision were assessed by adding analytes to levels of 0.5, 1, and 1.5 times the permissible maximum limit allowed in Brazil. The method presented selectivity with a decision limit (CC?) and detection capability (CC?) ranging from 114.2 to 143.7 and from 129.3 to 188.7 µg kg?1, respectively. The recovery of tetracyclines was higher than 82.5% with a precision of 7.1%, demonstrating theefficiency in determining tetracycline residues in cow milk.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call