Abstract

The use of an amplitude-modulated radiofrequency (RF) pulse for a control experiment is a proven method to control for off-resonance effects in multislice continuous arterial spin labeling (CASL) experiments. This method is also known as double adiabatic inversion. The adiabaticity factor of a single half-pulse, beta(1/2), and a new dimensionless parameter alpha, which is obtained from the labeling parameters and the flow velocity, are introduced. This makes it possible to distinguish three distinct cases: 1) With low alpha, a double inversion occurs. 2) With alpha > or = approximately 4, the efficiency with which the longitudinal magnetization is returned to the z-axis depends on the phase of the amplitude modulation at the time the spins cross the center of the labeling plane. 3) In the intermediate region, the efficiency shows undesirable fluctuations. In a Bloch equation simulation, three optimized parameter sets are determined. Near ideal performance should always be achieved by combinations of parameters for which beta(1/2) > or = approximately 2 and alpha approximately pi/beta(1/2). The efficiency increases were realized in a volunteer study, showing the practical application of the suggested optimization.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call