Abstract

Under ideal conditions, continuous arterial spin labeling (ASL) techniques are higher in SNR than pulsed ASL techniques by a factor of e. Presented here is a direct theoretical and experimental comparison of continuous ASL and pulsed ASL, using versions of both that are amenable to multislice imaging and insensitive to variations in transit times (continuous ASL with a delay before imaging, and QUIPSS II (Quantitative Imaging of Perfusion Using a Single Subtraction-second version)). Perfusion image quality for comparable imaging time was nearly identical for both single-slice and multislice imaging. The measured raw signal was approximately 25% higher with continuous ASL, but the SNR per unit time was identical.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.