Abstract
Omnidirectional, supercardioid, and adaptive directional microphones (ADM) were evaluated in combination with the ADRO amplification scheme for eight participants with moderate sloping hearing losses. The ADM produced better speech perception scores than the other two microphones in all noise conditions. Participants performed the Hearing in Noise Test sentences at -4.5 dB SNR or better, which is similar to the level achievable with normal hearing. The Speech, Spatial and Qualities of Hearing Scale indicated no disadvantages of using the ADM relative to the omnidirectional microphone in real-life situations. The ADM was preferred over the omnidirectional microphone in 54% of situations, compared to 17% preferences for the omnidirectional microphone, and 29% no preference. The combination of the ADM to improve SNR, and ADRO to keep the signal output comfortable and audible provided near-normal hearing performance for people with moderate hearing loss. The ADM is the recommended microphone configuration for ADRO hearing aids.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.