Abstract

A Chrompack PLOT fused-silica capillary column with a narrow internal diameter (0.25 mm) and PoraPLOT Q stationary phase are used for the simultaneous gas chromatographic—mass spectrometric determination of nitrogen, oxygen, carbon monoxide, carbon dioxide, nitrous oxide, nitric oxide, hydrogen cyanide, cyanogen, and water from thermal decompositions of the highly energetic materials 1,3,3-trinitroazetidine (TNAZ) and 5-nitro-2,4-dihydro-3H-1,2,4-triazol-3-one. Previously this laboratory required two separate capillary columns for analyses of nitrogen, oxygen, carbon monoxide, carbon dioxide, and nitrous oxide. The detection of hydrogen cyanide and cyanogen gases as products of TNAZ thermal decomposition is verified for the first time by our laboratory using this new column.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.