Abstract

A microelectrode array (MEA) is a configuration of multiple electrodes that enables the concurrent targeting of multiple sites for extracellular recording and stimulation. By utilizing light pulses or electrical stimulations, MEA recordings unveil the complex patterns of electrical activities that arise from the signaling processes within the retinal network. Here, we present a stepwise approach for using microelectrode arrays (MEAs) for recording action potentials from the mouse retina in response to electrical and light stimuli. We provide detailed techniques accompanied by description of a custom optical system, example recordings, troubleshooting guidelines, and data processing methods including spike sorting and code resources for analyzing light and electrical responses. The comprehensive nature of this paper aims to guide researchers in utilizing MEAs effectively for investigating retinal functionality. In particular, it can be easy to have a MEA experiment fail, but hard to identify the source of the failure. This paper is meant to demystify that process. It includes:•A description of MEA setup, recording, and spike data validation.•A troubleshooting guide for common failure modes in MEA recordings from mouse retina.•Spike detection and sorting to precisely extract distinctive action potential.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call