Abstract

A technique to correct the time walk due to the rise-time change has been developed in order to improve the time resolution. of coaxial Ge(Li) detectors. Typical time resolutions (fwhm) of 2.0 ns for 1332 keV gamma rays and 3.4 ns for 511 keV gamma rays have been obtained for a 60 cm 3 coaxial Ge(Li) detector. It is emphasized that the improvement can be achieved without any loss of efficiency.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.