Abstract
We develop an x-ray Ti/Au transition-edge sensor (TES) with an Au absorber deposited on the center of TES and improved its energy resolution using the K-means clustering algorithm in combination with Wiener filter. We firstly extract the main parameters of each recorded pulse trace, which are adopted to classify these traces into several clusters in the K-means clustering algorithm. Then real traces are selected for energy resolution analysis. Following the baseline correction, the Wiener filter is used to improve the signal-to-noise ratio. Although the silicon underneath the TES has not been etched to reduce the thermal conductance, the energy resolution of the developed x-ray TES is improved from 94 eV to 44 eV at 5.9 keV.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.