Abstract

Abstract Previous studies have demonstrated that mobile phones can be used as suitable emergency dosimeters in case of an accidental radiation overexposure. Glass samples extracted from displays of mobile phones are sensitive to ionizing radiation and can be measured using the thermoluminescence (TL) method. A non-radiation induced background signal (so-called zero dose signal) was observed which overlaps with the radiation induced signal and consequently limits the minimum detectable dose. Investigations of several glasses from different displays showed that it is possible to reduce the zero dose signal up to 90% by etching the glass surface with concentrated hydrofluoric acid. With this approach a reduction of the detection limit of a factor of four, corresponding to approximately 80 mGy, was achieved. Dosimetric properties of etched samples are presented and developed protocols validated by dose recovery tests under realistic conditions. With the improvements in sample preparation the proposed method of dose determination is a competitive alternative to OSL/TL measurements of electronic components and chip cards and provides a useful option for retrospective accident dosimetry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.