Abstract

In this paper, we have revisited a well‐known issue in quantitative X‐ray photoelectron spectroscopy (XPS) analysis: when quantitative XPS analysis is based solely on measured peak intensities, this can give rather confusing and misleading results. The problem is caused by peak intensity attenuation with photoelectron distance traveled in the sample and is an issue for samples where the atomic concentration varies substantially over the outermost ~5 nm. This is however the most significant depth range for many of the practical applications of nano‐structures which is today of increasing technological importance. We discuss 3 different methods to correct for the attenuation effect and their limitations. It is then pointed out how one can, by visually comparing observed peak shapes to a set of model spectra, often get a rough idea of the atom depth distribution in the surface region of the sample. This provides a practical tool to supplement the information obtained with XPS analysis based on peak intensities and chemical state information obtained from high‐resolution spectra.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.