Abstract

BackgroundSuccessful osseointegration of joint replacement implants is required for long-term implant survival. Accurate assessment of osseointegration could enable clinical discrimination of failed implants from other sources of pain avoiding unnecessary surgeries. Photon-counting detector computed tomography (PCD-CT) provides improvements in image resolution compared to conventional energy-integrating detector CT (EID-CT), possibly allowing better visualization of bone-implant-interfaces and osseointegration. The aim of this study was to assess the quality of visualization of bone-implant-interfaces and osseointegration in acetabular cup implants, using PCD-CT compared with EID-CT.MethodsTwo acetabular implants (one cemented, one uncemented) retrieved during revision surgery were scanned using PCD-CT and EID-CT at equal radiation dose. Images were reconstructed using different reconstruction kernels and iterative strengths. Delineation of the bone-implant and bone-cement-interface as an indicator of osseointegration was scored subjectively for image quality by four radiologists on a Likert scale and assessed quantitatively.ResultsDelineation of bone-implant and bone-cement-interfaces was better with PCD-CT compared with EID-CT (p ≤ 0.030). The highest ratings were given for PCD-CT at sharper kernels for the cemented cup (PCD-CT, median 5, interquartile range 4.25–5.00 versus EID-CT, 3, 2.00–3.75, p < 0.001) and the uncemented cup (5, 4.00–5.00 versus 2, 2–2, respectively, p < 0.001). The bone-implant-interface was 35–42% sharper and the bone-cement-interface was 28–43% sharper with PCD-CT compared with EID-CT, depending on the reconstruction kernel.ConclusionsPCD-CT might enable a more accurate assessment of osseointegration of orthopedic joint replacement implants.Key points• The bone-implant interface ex vivo showed superior visualization using photon-counting detector computed tomography (PCD-CT) compared to energy-integrating detector computed tomography.• Harder reconstruction kernels in PCD-CT provide sharper images with lower noise levels.• These improvements in imaging might make it possible to visualize osseointegration in vivo.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call