We compared the results of 505 urine specimens prospectively processed by both conventional manual processing (MP) with 16–24h incubation to BD Kiestra™ Total Laboratory Automation (TLA) system with a shortened incubation of 14h: 97% of culture results were clinically concordant. TLA processing was associated with improved standardization of time of first culture reading and total incubation time.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.