Abstract

Methods for long-term recording of peripheral nerve activity via intrafascicular electrodes have not been optimized. We compared the long-term functionality of custom-made 95%Pt/5%Ir intrafascicular electrodes containing a proximal spring-like structure to that of conventional straight electrodes. The modified electrode was implanted into the sciatic nerve fascicle of a random hind limb in 14 rabbits for 9 months. A conventional electrode was implanted in the opposite hind limb as a control. Orthodromic and antidromic nerve potentials were sampled and analyzed monthly. Latency, amplitude, and nerve conduction velocity of electrical signals were generally similar within the modified group and straight control group at different time intervals (P > 0.05). However, at the conclusion of the study period, the modified electrode group had a greater number of functioning electrodes (P < 0.05) and a greater total functioning electrode time (P = 0.006). Intrafascicular electrodes with a spring-like structure demonstrated superior potential for long-term electrophysiological monitoring over straight electrodes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.