Abstract

The fast neutron energy dependence of response of the HPA PADC neutron personal dosemeter has been measured from 144 keV to 19 MeV using monoenergetic neutron fields. Below 144 keV the relative energy and angle dependence of response have been determined using MCNP-4C2. New data from the SIGMA field at Cadarache, France, have been used to determine the appropriate scaling factor for the calculated response to thermal and intermediate energy neutrons. These newly determined response characteristics of the dosemeter are discussed with respect to its performance in the EVIDOS workplace field irradiations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.