Abstract
Echo planar imaging (EPI) is an ultrafast magnetic resonance imaging (MRI) technique that allows one to acquire a 2D image in about 100 ms. Unfortunately, the standard EPI images suffer from substantial geometric distortions, mainly originating from susceptibility differences in adjacent tissues. To reduce EPI distortions, correction methods based on a field map, which is a map of the off-resonance frequencies, have been developed. In this work, a nonlinear least squares estimator is used to optimize the estimation of the field map of the B(0) field. The model of the EPI and reference data includes parameters for the phase evolution, the complex magnitude, the relaxation of the MRI signal and the EPI-specific phase difference between odd and even echoes, and from these parameters, additional corrections might be computed. The reference data required to estimate the field map can be acquired with a modified EPI-sequence. The proposed method is tested on simulated as well as experimental data and proves to be significantly more robust against noise, compared to the previously suggested method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.