Abstract
We have investigated the effects of sample rotation and ion energy reduction on the quality of ion sputtering assisted Auger electron spectroscopy depth profiling of W/WNx and WSix films on polysilicon substrates. Rotation significantly improves both the depth resolution for W/WNx/Si samples and the sensitivity to oxygen inside the W/WNx film. Rotation does not have any effect on results for WSix/Si interfaces. Atomic force microscopy investigations show that the improved resolution for W/WNx/Si samples is due to reduced roughness of the sputtered surface when the sample is rotated. Reduction of the ion energy from 4 to 1 keV also improves the depth profiles of W/WNx films. The ion energy reduction in the same range leads to larger error in measured elemental concentrations for WSix films due to preferential sputtering. This reduction increases the depth resolution of Auger depth profiles in the case of polycrystalline WSix films, but does not change the depth resolution for amorphous WSix films.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Vacuum Science & Technology A: Vacuum, Surfaces, and Films
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.