Abstract

The CIEMAT/NIST efficiency tracing method has been widely used for radionuclide standardization. The method consists of the theoretical calculations of counting efficiencies and the experimental data obtained from a commercially available double photomultiplier detector. We have recently implemented the technique using 3H as a tracer. Our implementation of the method was validated by participating into an international comparison of activity measurements of a solution of 204T1. We report in this paper the procedure employed to implement the technique and compare our measurement result of 204T1 activity with that of the international comparison.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.