Abstract

Trunnionosis is emerging as an early mode of failure in conventional metal-on-polyethylene total hip arthroplasty. It is defined as wear or corrosion at the trunnion, the taper at the femoral head-neck interface. Trunnion wear can result in a variety of negative sequelae and, in severe cases, necessitate revision arthroplasty. We describe a 64-year-old man with a metal-on-polyethylene total hip arthroplasty who presented with a sensation of clunking in the hip. Initial imaging and laboratory studies were inconclusive, and the decision was made to monitor. Two years later, trunnion wear was detected on radiographs, presenting as an abnormal alignment of the femoral neck relative to the femoral head. Several case reports and series describe catastrophic total hip arthroplasty failure due to trunnionosis. However, few describe the radiographic signs of wear at the trunnion before gross failure. This early presentation is important to recognize to minimize patient morbidity and aid surgical planning.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.