Abstract
We describe a case of total hip arthroplasty failure where a cobalt-chrome femoral head completely wore through the polyethylene liner and the titanium acetabular cup. The patient subsequently underwent revision total hip arthroplasty with acetabular revision and femoral head exchange. This case illustrates the natural history of catastrophic failure in a metal-on-polyethylene total hip arthroplasty design. If recognized earlier, this patient may have been a candidate for isolated liner and head exchange. Additionally, this case represents a unique complication in using mixed metals in total hip arthroplasty, where the harder cobalt-chrome femoral head wore completely through the much softer titanium component.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.