Abstract

To measure non-invasively the impedance index of a range of geometries of PTFE end-to-side anastomoses (ESA). In vitro experiments using a custom-built flow rig. Anastomoses constructed at each angle. (15 degrees, 30 degrees, 45 degrees, 60 degrees, 90 degrees, 120 degrees) for both standard and cuffed ESA. The impedance index of each ESA increased non-linearly with flow rate. The impedance index at a given flow rate was reduced by decreasing the anastomotic angle and further reduced by the addition of a Miller cuff. The reduction in impedance index achieved with a Miller cuff may help to explain the improved patency rates of femoropopliteal grafts incorporating a cuff.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.