Abstract

Impedance is the ratio of voltage to current in an electrical circuit. Cardiovascular implantable electronic devices measure impedance to assess the structural integrity electrical performance of leads, typically using subthreshold pulses. We review determinants of impedance, how it is measured, variation in clinically measured pacing and high-voltage impedance and impedance trends as a diagnostic for lead failure and lead-device connection problems. We consider the differential diagnosis of abnormal impedance and the approach to the challenging problem of a single, abnormal impedance measurement. Present impedance provides a specific but insensitive diagnostic. For pacing circuits, we review the complementary roles of impedance and more sensitive oversensing diagnostics. Shock circuits lack a sensitive diagnostic. This deficiency is particularly important for insulation breaches, which may go undetected and present with short circuits during therapeutic shocks. We consider new methods for measuring impedance that may increase sensitivity for insulation breaches.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.