Abstract

Images were obtained with a cassette spot film device with an inserted portable flat-panel detector (Fujifilm, Tokyo, Japan). The CDRAD phantom was sandwiched between polymethylmethacrylate (PMMA) with total thicknesses of 12, 16, 21, and 26 cm to simulate patient attenuation and scatter. Tube voltages of 81, 90, 109, and 125 kVp were used to make the radiographs. In total, 12 different Virtual Grid settings (grid ratio, line pairs (lp)/cm, and type of interstitial material) were applied for every phantom thickness and tube voltage. An average increase of 32% in IQ was obtained when Virtual Grid images with a SW grid ratio 10:1 were compared with gridless images (P < 0.001). Increasing the SW grid ratio to 20:1 resulted in a further increased IQ. With a phantom thickness of 12 cm PMMA, Virtual Grid outperformed the removable physical grid presented in the cassette spot film device. The linear mixed-effects model showed that IQ is mainly affected by PMMA, tube voltage, and the SW grid ratio. Virtual Grid improves images obtained without physical grid for a wide range of experimental conditions. Despite the different possible settings of the Virtual Grid SW, the most important parameter affecting IQ is the SW grid ratio.