Abstract

Needle-free connectors (NFCs) were introduced to eliminate the use of needles in intravascular catheters, and their newest generations were designed to improve patient safety and reduce catheter-related bloodstream infection (CRBSI) risks. The aim of this meta-analysis was to compare NFCs with 3-way stopcocks (3WSCs) and their effects on CRBSI rates. A meta-analysis was conducted using a research protocol consistent with the PRISMA statement for reporting meta-analyses. The Cochrane Database of Systematic Reviews and MEDLINE were searched for relevant randomized studies published from January 2000 to September 2018. We identified and selected for the meta-analysis 8 studies comparing CRBSI rates (according to the Centers for Disease Control and Prevention's National Healthcare Safety Network definition) associated with NFCs utilizing negative-displacement, neutral-displacement, or positive-displacement devices with rates for 3WSCs. Relative risk was 0.53 with a 95% CI of 0.28 to 1.00, and the relative difference was -0.018 with a 95% CI of -0.039 to 0.004. CRBSI risk was statistically higher for 3WSCs compared to NFCs.