Abstract

PurposeThe objective of this study was to evaluate the impact of the magnetic field regulation in conjunction with the volumetric repainting technique on the spot positions and range in pencil beam scanning proton therapy.Methods“Field regulation” — a feature to reduce the switching time between layers by applying a magnetic field setpoint (instead of a current setpoint) has been implemented on the proton beam delivery system at the Miami Cancer Institute. To investigate the impact of field regulation for the volumetric repainting technique, several spot maps were generated with beam delivery sequence in both directions, that is, irradiating from the deepest layer to the most proximal layer (“down” direction) as well as irradiating from the most proximal layer to the deepest layer (“up” direction). Range measurements were performed using a multi‐layer ionization chamber array. Spot positions were measured using two‐dimensional and three‐dimensional scintillation detectors. For range and central‐axis spot position, spot maps were delivered for energies ranging from 70–225 MeV. For off‐axis spot positions, the maps were delivered for high‐, medium, and low‐energies at eight different gantry angles. The results were then compared between the “up” and “down” directions.ResultsThe average difference in range for given energy between “up” and “down” directions was 0.0 ± 0.1 mm. The off‐axis spot position results showed that 846/864 of the spots were within ±1 mm, and all off‐axis spot positions were within ±1.2 mm. For spots (n = 126) at the isocenter, the evaluation between “up” and “down” directions for given energy showed the spot position difference within ±0.25 mm. At the nozzle entrance, the average differences in X and Y positions for given energy were 0.0 ± 0.2 mm and −0.0 ± 0.4 mm, respectively. At the nozzle exit, the average differences in X and Y positions for given energy were 0.0 ± 0.1 mm and −0.1 ± 0.1 mm, respectively.ConclusionThe volumetric repainting technique in magnetic field regulation mode resulted in acceptable spot position and range differences for our beam delivery system. The range differences were found to be within ±1 mm (TG224). For the spot positions (TG224: ±1 mm), the central axis measurements were within ±1 mm, whereas for the off‐axis measurements, 97.9% of the spots were within ±1 mm, and all spots were within ±1.2 mm.

Highlights

  • Pencil beam scanning (PBS) delivery technique has become a preferred method relative to passive scattering methods in proton therapy.[1,2] Pencil beam scanning technique delivers a single pristine beam at a time

  • | 125 whereas for the off‐axis measurements, 97.9% of the spots were within ±1 mm, and all spots were within ±1.2 mm

  • The off‐axis spot position results demonstrated that 97.9% (846/864) of the spots were within ±1 mm, and all off‐axis spot positions were within ±1.2 mm

Read more

Summary

Introduction

Pencil beam scanning (PBS) delivery technique has become a preferred method relative to passive scattering methods in proton therapy.[1,2] Pencil beam scanning technique delivers a single pristine beam at a time. The whole target volume is irradiated by delivering the proton beam either from the deepest layer to the most proximal layer (“down” direction) or from the most proximal layer to the deepest layer (“up” direction), or a combination of both. The repeated scans of the entire volume allow the delivery of planned dose to the tumor volume repeatedly, providing the statistical averaging of dose heterogeneity.[14]

Objectives
Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.