Abstract
BackgroundThe aims of this study were (1) to evaluate the impact of injection dose, post-reconstruction filtering, and collimator choice on image quality of myocardial perfusion single-photon emission computed tomography (SPECT) using cadmium-zinc telluride (CZT) detectors and (2) to determine how these factors affect measured infarct size in the in vivo rat.MethodsTwenty-four healthy and eight myocardial infarct (MI) rats underwent myocardial perfusion SPECT imaging after injection of various doses (25 to 200 MBq) of 99mTc-tetrofosmin using a standard (STD) five-pinhole collimator and high-sensitivity (HS) five-pinhole collimator. Image quality score, contrast-to-noise ratio, sharpness index, coefficient of variation (CV), and measured defect size were assessed as measures of image quality.ResultsThe image quality score increased and CV decreased as a function of injection dose. The contrast-to-noise ratio increased and sharpness index decreased as a function of Gaussian kernel size. When STD and HS were compared, HS tended to show higher image quality score and lower CV than STD. The use of post-reconstruction filter significantly improved image quality score and lessened CV. The reproducibility of defect size measurements, as assessed by intraclass correlation coefficients (ICC), between the collimators was poor-to-moderate (ICC = −0.31~0.57) with low (25 MBq) injection dose and with no or light (1.5-mm kernel size) filtering, whereas it was good-to-excellent (ICC = 0.75~0.97) with high (200 MBq) dose or low dose with heavy (2.5-mm kernel size) filtering. The filtering-related reproducibility was poor (ICC = −0.18~0.17) for STD with low injection dose, whereas it was good-to-excellent (ICC = 0.79~0.89) for HS. Furthermore, there was a filtering-related underestimation of defect size particularly with the use of heavy smoothing.ConclusionsAppropriate imaging setting is important to obtain high quality images and thereby reliable measurements using a preclinical myocardial SPECT in the rat. When only a low injection dose (25 MBq) is allowed, we would recommend to use HS with light (1.5-mm kernel size) filtering.
Highlights
The aims of this study were (1) to evaluate the impact of injection dose, post-reconstruction filtering, and collimator choice on image quality of myocardial perfusion single-photon emission computed tomography (SPECT) using cadmium-zinc telluride (CZT) detectors and (2) to determine how these factors affect measured infarct size in the in vivo rat
Without post-reconstruction filtering, the image quality was poor when the injection dose was low (e.g., 25 MBq) with STD, which improved as a function of injection dose
When STD and HS were compared, the image quality of HS tended to be higher than STD, which was statistically significant at the injection dose of 50 and 200 MBq without post-reconstruction filtering
Summary
The aims of this study were (1) to evaluate the impact of injection dose, post-reconstruction filtering, and collimator choice on image quality of myocardial perfusion single-photon emission computed tomography (SPECT) using cadmium-zinc telluride (CZT) detectors and (2) to determine how these factors affect measured infarct size in the in vivo rat. Preclinical molecular imaging such as single-photon emission computed tomography (SPECT) and positron emission tomography (PET) using small animals is increasingly being recognized as an important tool for cardiovascular research [1,2]. Such data would be necessary for reliable SPECT measurements before a biomedical question can be addressed
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have