Abstract

To compare radiation doses given to patients undergoing IVU (intravenous urography) before and after digitalization of our X-ray department. IVU examinations were monitored with dose area product meters before and after the X-ray department changed to digital techniques. The first step was a change from film-screen to storage phosphor plates, while the second step involved changing to a flat panel detector. Forty-two patients were included for the film-screen situation, 69 when using the storage phosphor plates, and 70 using the flat panel detector. A dose reduction from 41.8 Gycm2 to 31.5 Gycm2 was achieved with the first step when the film-screen system was replaced with storage phosphor plates. A further reduction to 12.1 Gycm2 was achieved using the flat panel detector. The introduction of the flat panel detectors made a considerable dose reduction possible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.