Abstract

ObjectivesMost of cardiac dedicated CZT-SPECT systems are not equipped with CT, whereas PET systems are. We evaluated the impact of AC correction on CZT-SPECT myocardial blood flow (MBF) and myocardial flow reserve (MFR) measurements. Methods104 patients were included. SPECT data were acquired on cadmium zinc telluride (CZT)-based pinhole cardiac camera in listmode using a stress (250 ± 17 MBq)/rest (511 ± 23 MBq) 1-day Tc-99m-tetrofosmin protocol. Low-dose CT was acquired on another SPECT/CT camera in the same position. All analysis was performed using Corridor4DM. ResultsStress and rest MBF were significantly lower when AC was applied (P < 0.001). For regional and global MFR, there was no significant difference between AC and NAC measurements (P > 0.25 at least). Mean global LV MFR was 2.43 ± 0.87 and 2.33 ± 0.89, respectively, for NAC and AC measurements. Using a threshold of 2, 86 patients (83%) remained classified as normal and abnormal regarding global LV MFR whether AC was applied or not. Mean difference between NAC and AC values for the 18 other patients was 0.3. ConclusionAC correction does not significantly affect MFR measurement both in regional and global LV analyses.