Abstract

A sensitive, highly specific immunoassay method has been developed by measuring the enhanced resonance light scattering (RLS) signals of immunoreactions with simultaneously scanning both the excitation and the emission monochromators of a common spectrofluorometer. For a given content of antibody (Ab), the RLS signals of an immunoreaction follow Gaussian distribution with antigen (Ag) concentration. The central position of the Gaussian curve represents the concentration of given Ab, and the half bandwidth has proved to be a characteristic constant of a given Ab–Ag immunoreaction. With the RLS signals, the limit of detection for human immunoglobulin G (HIgG) in serum samples could reach 10 ng ml −1, and the concentration of HIgG in blood serum samples could be detected with the recovery of 90.2–107.7% and R.S.D. of 0.8–2.7%. The results of determination for three human serum samples are identical to those obtained by immunoturbidimetry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.