Abstract

ObjectivesTo compare image quality and radiation dose of right adrenal vein (RAV) imaging computed tomography (CT) among conventional, low kV, and low kV with reduced contrast medium protocols. MethodsOne-hundred-and-twenty patients undergoing adrenal CT were randomly assigned to one of three protocols: contrast dose of 600mgI/kg at 120-kV tube voltage setting (600-120 group), 600mgI/kg at 80kV (600-80 group), and 360mgI/kg at 80kV (360-80 group). Iterative reconstruction was used for 80-kV groups. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of the RAV and size-specific dose estimates (SSDE) were measured. Three radiologists evaluated 4-point visualisation scores of RAV by consensus reading. ResultsThe RAV detectability was 95%, 97.2%, and 97.3% for 600-120, 600-80, and 360-80 groups, respectively (p=1.000). Visualisation scores were not significantly different among the groups (p=0.152). There were no significant differences in CNR or SNR between the 600-120 and 360-80 groups. SSDE of the 360-80 group was significantly lower than that of the 600-120 group (5.86mGy±1.44 vs. 7.27mGy±1.81, p<0.001). Conclusions80-kV scans with 360 mgI/kg contrast media showed comparable detectability of RAV to conventional scans, while reducing 19% of SSDE.