Abstract

A simple model structure of spatial and temporal temperature distribution in human tissue, exposed to electromagnetic heating and surface cooling, is developed. This model, denoted the control model, is useful for thermal dosimetry in hyperthermia cancer therapy. The theoretical control model parameter values for various types of homogeneous tissue are determined from a priori knowledge about tissue composition and parameters. The control model parameters for real tissue are estimated from patient treatment data through the use of a system identification technique. The results indicate that the control model structure is adequate. With the available a priori knowledge the theoretically determined control model parameter values do not accurately reproduce the experimentally estimated values. Accordingly, an identification based on experimental data is recommended, if an accurate model is required.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.