Abstract

A simple, inexpensive optical mount for use with the Atomsource glow discharge direct solids atomiser has been designed and built. This mount allows adjustment of the atomiser in the horizontal, vertical and rotational (in the horizontal plane) modes relative to the optical path. The mount has been used in conjunction with the associated spectrometer, the Analyte 16 Atomic Absorption Spectrometer, to illustrate the presence of concentration gradients of sputtered species within the Atomsource plasma. From these data, general mass transport patterns and regions of enhanced ionisation in the Atomsource can be identified. The optimum location in the plasma from which to sample using atomic absorption detection of sputtered material can also be established.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.