Abstract
A power splitter, a delay line, a dual-channel 16-bit pipeline analogue-to-digital converter (ADC), and a low-cost clock synthesiser are used with digital signal processing techniques to measure cycle-to-cycle random jitter. The proposed setup is capable of measuring the jitter of a device under test (DUT), using an ADC clock that has an order of magnitude higher jitter than the DUT. Experimental results show that the average jitter reading of a low phase noise frequency synthesiser is 312 fs rms, with a 38 fs standard deviation between multiple readings. An imprecise clock source with 7.7 ps rms jitter is used to clock the ADCs to do this 312 fs measurement. The method is compared to a 33 GHz oscilloscope, a low-noise spectrum analyser, and a high performance signal source analyser (SSA). The results are much more accurate than both the oscilloscope and the spectrum analyser, and are reasonably close to the results of the SSA.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.