Abstract
Animal and human experiments using a multi-channel EIT system requires a cumbersome procedure to attach multiple electrodes. We have to ensure good contact of all electrodes and manage many lead wires during experiments. The problem becomes more severe as we increase the number of electrodes. These may limit the applicability of the imaging method in practice. Noting this technical difficulty, there have been a few trials to design human interface means such as electrode belts, helmets or rings. In this study, we developed an electrode belt for long-term monitoring of human lung ventilation. The belt includes 16 embossed electrodes which make good contact with the skin. The electrode is made by conductive polymer and metallic thread. Soft cushion and wide contact area minimize uncomfortable sensation and reduce contact impedances. The electrodes are attached to an elastic fabric belt at equal spacing. We describe details of its design and fabrication. Using the electrode belt and recently developed multi-frequency EIT system KHU Mark2, we show time-difference chest images of three human subjects during normal breathing cycles.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have