Abstract

A simple, rapid, and reproducible method for the determination of urea, creatinine, and uric acid in human serum and urine by ion-pair reversed-phase high-performance liquid chromatography with UV detection (196 nm) has been developed. The method involves the pretreatment of serum samples with trichloroacetic acid and centrifugation followed by the isocratic separation of compounds on a μ-Bondapak C18 column using a mobile phase consisting of 1.25 mmol/l tetrabutylammonium phosphate. For urine samples no special pretreatment is necessary. In addition, this method allows, with some limitations, the determination of creatine in serum.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.