Abstract

We describe a convenient method for the separation and quantification of xanthine, hypoxanthine, and uric acid in 20 microL of cerebrospinal fluid (CSF) with use of HPLC and ultraviolet detection. The analysis is performed on a Sepharon SGX C18 column and the elution system consists of potassium phosphate buffer, pH 5.1, with 20 mL/L methanol. The lower limit of detection was 4 pmol for hypoxanthine and xanthine and 6 pmol for uric acid. Analytical recoveries of purine metabolites ranged from 98.6% to 102.9%. The intra- and interassay CVs were <3%. The applicability of the method is illustrated with the determination of micromolar concentrations of xanthine, hypoxanthine, and uric acid in CSF samples obtained from 113 patients with various neurological disorders.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.