Abstract

High-performance liquid chromatography (HPLC) with UV, circular dichroism (CD) and intrinsic fluorescence detection was applied to monitor conformational properties of recombinant human interferon α2b when performing size exclusion chromatography (SEC) and reversed-phase HPLC (RP-HPLC). In this way native conditions during SEC and structural changes of the protein during RP-HPLC were demonstrated. These results were confirmed by stand-alone fluorescence and CD measurements. With respect to HPLC tandem detection, the fluorescence detector compared favourably to the UV and CD detector regarding linearity, sensitivity and selectivity. SEC combined with intrinsic fluorescence scanning detection permits conformational analysis of small amounts of aggregates in the presence of excess native monomeric protein. In conclusion, HPLC with on-line UV and intrinsic fluorescence detection provides a promising concept for analysing the amount and conformational properties of a biopharmaceutical and its impurities.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.