Abstract

Both polymer gels and PRESAGE radiochromic solid dosimeter, in conjunction with optical CT scanning system, have been employed to measure 3-D dose distribution. The 3-D dose maps obtained from these systems can provide a useful tool for dose verification on complex treatments such as IMRT, radiosurgery, and RapidArc. These complex treatments present high dose gradient regions in the boundaries between the target and the surrounding critical organs. Dose accuracy in these areas can be critical, and may affect the treatment. There is a pressing need for a dosimeter that allows for accurate determination of 3-D dose distribution with high spatial resolution. In this paper, performance of polymer gels and PRESAGE dosimeter with optical CT scanning is reviewed and evaluated in terms of their sensitivity calibration, irradiation, optimization of scanning procedures, precision, and accuracy. Clinical applications of optical-CT dosimetry are presented.