Abstract

Transportable (e.g., "giraffe") air samplers can be used for both short-term "grab-sampling" situations involving a few minutes to several hours, as well as for long-term retrospective air sampling situations that typically last a week. It is the short-term coverage situation which begs the question: "how long a sampling time is adequate?" The basis for the answer (which depends mainly on the selected volumetric sampling rate and the counter sensitivity, as well as the concentration level to be measured) can be simply viewed as follows: that an adequate activity must be measureable on the air filter sample at the cessation of sampling. Equations for selection of sampling times are presented, as are example calculations for alpha counters in order to measure various target concentrations and exposures. The concepts presented here can be readily adapted to other counting systems as well as to different sampling flow rates.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.