Abstract

Background and objective: Saline infusion is applied together with radiofrequency ablation (RFA) to enlarge the ablation zone. However, one of the issues with saline-infused RFA is backflow, which spreads saline along the insertion track. This raises the concern of not only thermally ablating the tissue within the backflow region, but also the loss of saline from the targeted tissue, which may affect the treatment efficacy. Methods: In the present study, 2D axisymmetric models were developed to investigate how saline backflow influence saline-infused RFA and whether the aforementioned concerns are warranted. Saline-infused RFA was described using the dual porosity-Joule heating model. The hydrodynamics of backflow was described using Poiseuille law by assuming the flow to be similar to that in a thin annulus. Backflow lengths of 3, 4.5, 6 and 9 cm were considered. Results: Results showed that there is no concern of thermally ablating the tissue in the backflow region. This is due to the Joule heating being inversely proportional to distance from the electrode to the fourth power. Results also indicated that larger backflow lengths led to larger growth of thermal damage along the backflow region and greater decrease in coagulation volume. Hence, backflow needs to be controlled to ensure an effective treatment of saline-infused RFA. Conclusions: There is no risk of ablating tissues around the needle insertion track due to backflow. Instead, the risk of underablation as a result of the loss of saline due to backflow was found to be of greater concern.