Abstract

We describe a versatile high-performance liquid-chromatographic method for determining homocysteine and other plasma sulfhydryls. Using three different procedures for preparation of plasma, we determined total, free (non-protein-bound), and reduced forms of homocysteine, cysteine, glutathione, cysteinylglycine, and gamma-glutamylcysteine in human plasma. Sample preparation involves disulfide reduction with dithiothreitol and protein precipitation with sulfosalicylic acid. The assay utilizes isocratic reversed-phase ion-pair liquid chromatography at pH 2.4, postcolumn derivatization with 4,4'-dithiodipyridine, and colorimetric detection at 324 nm. The intra-assay precision (CV) of the method for total homocysteine is 1.5%; the interassay precision over 2.5 months is 2.5%. The detection limit for homocysteine is < 50 nmol/L plasma.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.