Abstract
The aim of this study is to visualize the coronary artery with the intravenus injection of contrast media. We propose a method of rapid and successive acquisition of two-dimensional images in a digital processing system which uses a pair of video cameras, a shutter operation, and a beam-splitting apparatus. A prototype of the high-speed acquisition system was constructed. The two-monochromatic video-camera method was very suitable for quick acquisition. Electro-optical shutters (PLZT) were found practical with respect to the dynamic range and response time. The field blanking method was essential for the data acquisition.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.