Abstract

High-performance size-exclusion chromatography of some standard proteins, peptides and amino acids on a new hydrophilic packing material obtained by chemical transformation of a cross-linked polystyrene-divinylbenzene copolymer was studied. Columns filled with 4 and 7 μm particles were compard. The influence of the concentration of acetonitrile, isopropanol and trifluoroacetic acid in the mobile phase on the chromatographic performance was investigated. A good linear calibration graph covering the molecular weight range from 200 to 700 000, was obtained under the optimal conditions. The packing material can be used for separations, for molecular weight determinations and for the pre-fractionation of proteins. The high rigidity of the packing material allows relatively high pressures to be used and therefore fast separations to be achieved. The packing material was applied to the chromatography of proteins from beer, bones and milk.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call