Abstract

Reported is an lateral-mode resonant cantilever sensor that features a much depressed liquid drag force, thereby, achieving a much higher quality factor of 249 in water and higher specific mass sensitivity in liquids, compared with traditional out-of-plane resonant cantilevers. With 8.8Hz/pg sensitivity calibrated by specific binding streptavidin-labled Dynabeads with the biotin-immobilized cantilever in Phosphate Buffered Saline (PBS) solution, specific-antibody immobilized cantilevers have real-time detected E. coli in PBS solution. Real-time detection of Hg2+ ions in aqueous Hg(NO 3 ) 2 is also carried out by using the sensing-group-modified cantilevers, resulting in a finer noise-limited resolution than 100ppb.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.