Abstract

A sensor for simultaneous measurements of temperature and carbon dioxide (CO2) concentration at elevated pressure is developed using tunable diode lasers at 2 µm. Based on some selection rules, a CO2 line pair at 5006.140 and 5010.725 cm(-1) is selected for the TDL sensor. In order to ensure the accuracy and rapidity of the sensor, a quasi-fixed-wavelength WMS is employed. Normalization of the 2f signal with the 1f signal magnitude is used to remove the need for calibration and correct for transmission variation due to beam steering, mechanical misalignments, soot, and windows fouling. Temperatures are obtained from comparison of the background-subtracted 1f-normalized WMS-2f signals ratio and a 1f-normalized WMS-2f peak values ratio model. CO2 concentration is inferred from the 1f-normalized WMS-2f peak values of the CO2 transition at 5006.140 cm(-1). Measurements of temperature and CO2 concentration are carried out in static cell experiments (P = 1-10 atm, T = 500-1200 K) to validate the accuracy and ability of the sensor. The results show that accuracy of the sensor for temperature and CO2 concentration are 1.66% temperature and 3.1%, respectively. All the measurements show the potential utility of the sensor for combustion diagnose at elevated pressure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.