Rapid and sensitive high-pressure liquid chromatographic (HPLC) assays for ticarcillin in serum and urine have been developed. Sample pretreatment and optimized chromatographic conditions are presented for a C-18-bonded reversed-phase column used in an internal standard assay method. Ticarcillin has a retention time of ~5.3min at a flow rate of 1.5ml/min for a mobile phase of acetonitrile-aqueous 0.06 M sodium biphosphate, pH 2.05, (50.5:100). In a two-step extraction procedure, the ticarcillin extraction efficiencies from serum and urine were 76.1 ± 4.7 and 80.9 ± 3.2%, respectively. The assay sensitivity limit for ticarcillin in these fluids is ~1.0 μg/ml. A comparison is made of the HPLC and microbiological assay results for ticarcillin in 20 different but equally divided serum samples obtained from two volunteers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.