Abstract
A glycopeptide antibiotic, teicoplanin, was used as chiral stationary phase for the high-performance liquid chromatographic (HPLC) separation of enantiomers of more than 30 unnatural amino acids, such as phenylalanine and tyrosine analogues and analogues containing 1,2,3,4-tetrahydroisoquinoline, tetraline, 1,2,3,4-tetrahydro-2-carboline, cyclopentane, cyclohexane, cyclohexene, bicyclo[2.2.1]heptane or heptene skeletons. Excellent resolutions were achieved for most of the investigated compounds by using a hydro-organic mobile-phase system. The effects of organic modifier content, temperature and flow-rate on the resolution were investigated and the conditions of separation were optimized.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.