Abstract

A method for the determination of total phenylephrine (free plus conjugated) in human serum was developed using column-switching high-performance liquid chromatography (HPLC) with fluorescence detection. After serum was deproteinized with acetonitrile, the conjugated phenylephrine was hydrolyzed with diluted hydrochloric acid. The solution was evaporated to dryness. The reconstituted residue was analyzed with HPLC using a heart-cut technique. Good recoveries of the analytes from spiked human serum samples were obtained with small coefficients of variation. A good linear response was obtained for the concentration range 5–500 ng/ml. The lower limit of quantitation for phenylephrine in human serum was 5 ng/ml. The method was applied to the determination of phenylephrine in human serum after oral administration of phenylephrine hydrochloride.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.