Abstract

A rapid procedure based on a direct extraction and HPLC determination of dihydroergocristine in a pharmaceutical preparation with fluorescence detection has been developed and validated. The optimized chromatographic conditions included a Purospher RP18e column, 5 microm particle size, 250 x 4.0 mm, and 25 mM potassium dihydrogen phosphate buffer (pH 2.8)-acetonitrile (60 + 40, v/v) mobile phase at a flow rate of 1 ml/min. The separation was carried out at 50 degrees C, and the injection volume was 5 microL. Fluorescence detection was performed at an excitation and emission wavelength of 224 and 344 nm, respectively. The mobile phase parameters such as organic solvent composition, temperature, and pH were studied. The proposed method has the advantages of a very simple sample pretreatment and fast HPLC determination.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.